Deterministic personal agent skill for calculating Pulmonary Functional Residual Capacity (FRC), Total Lung Capacity (TLC), Residual Volume (RV), and RV/TLC ratio.
README.md
Pulmonary Closed-Circuit Helium Dilution FRC Personal Agent Skill
Deterministic, zero-mock personal agent micro-skill for calculating Functional Residual Capacity (FRC), Total Lung Capacity (TLC), Residual Volume (RV), and RV/TLC ratio using closed-circuit helium dilution gas physics.
Architecture
flowchart TD
A[Initial Circuit C1 He & V_spiro] --> B[Closed Rebreathing & Equilibrium]
B --> C[Equilibrated Concentration C2]
C --> D[FRC Mass Balance Calculation]
D --> E[TLC & RV Stratification]
E --> F[Air Trapping vs Hyperinflation Alert]
Quickstart
from client import HeliumDilutionFRCClient
client = HeliumDilutionFRCClient()
result = client.evaluate_frc_volumes(
c1_initial_he_fraction=0.10,
c2_equilibrated_he_fraction=0.0667,
spirometer_circuit_volume_l=5.0,
inspiratory_capacity_l=3.0,
expiratory_reserve_volume_l=1.0
)
print(result)